Ionizing radiation metrology at the service of health: quality control of radiopharmaceutical dose calibrators in nuclear medicine unit -accurarcy, reproducibility and linearity tests-

20222 citationsOpen accessAbdelmalek Essaâdi University

Abstract

Ensuring the effective use of a dose calibrator used for the practice of nuclear medicine, and the injection of the correct dose to the patient is the subject of a long traceable calibration chain, which serves to establish the relationship between the values of the activity indicated by the dose calibrator and the activity corresponding to ISO 2919 certified sealed sources whose activity is known with an uncertainty of less than 1 %. Our calibration program consists in carrying out various quality control on the CRC 55TR activimeter, namely the daily control which is carried out in the absence of any radioactive source, then the fidelity control carried out by a source of 37 Cs and a source of 60 Co, then the reproducibility check performed with a 137Cs source, and finally the linearity check performed with a 99m Tc source. In our case the results obtained from the specific tests used to verify the proper functioning of the Capintec CRC 55tR activimeter in a nuclear medicine service in Morocco, shows an agreement with international standards so that the measurement of activities is done with significant results, in addition, to guarantee the metrological quality of the measurements delivered, Our laboratory registers in an accreditation process according to ISO 17025.

Research topics

  • Radioactive Decay and Measurement Techniques
  • Radiation Dose and Imaging
  • Radioactivity and Radon Measurements

Read the original research

This page summarises published work. The authoritative version sits with the publisher.

DOI: 10.1051/e3sconf/202235101095

Is something wrong with this record? Report it or request removal.

Discussion

Discuss this research

Have you built on this work, tried to replicate it, or seen it applied in practice? Share what you know. Verified researchers and MARATTO™ domain experts can open a discussion, and any member can reply. Contributions are reviewed before they appear.

No discussion yet. Open the first thread.